iCheck Zinc

iCheck™ Vitamin E is a state-of-the–art and the first of its kind portable fluorometer. It determines vitamin E as alpha tocopherol concentration in cattle whole blood and serum. iCheck™ Vitamin E measures the fluorescence in the test vial and calculates the vitamin E content in mg/L.

What is inside

iCheck Fluoro: Case Content

iCheck Fluoro Case Content
A

iCheck device

B

Rechargeable batteries & charger

C

USB -B cable for data transfer

D

Vial stand

E

Fluoro Standard

F

USB stick with User Manual

G

User Manual

H

Weighing dish

I

Calibration weight

J

Balance

A

iCheck device

B

Rechargeable batteries & charger

C

USB -B cable for data transfer

D

Vial stand

E

Fluoro Standard

F

USB stick with User Manual

G

User Manual

H

Weighing dish

I

Calibration weight

J

Balance

A – iCheck device, B – Rechargeable batteries & charger, C – USB -B cable for data transfer, D – Vial stand, E – Fluoro Standard, F – USB stick with User Manual, G – User Manual, H – Weighing dish, I – Calibration weight, J – Balance.

iCheck Fluoro: Test Kit Content

iCheck Fluoro Test Kit Contents
A

iCheck Fluoro Reagent Vials x 100

B

1 mL syringe x 100

C

Needle x 100

D

Vial labels/stickers

iCheck Fluoro Test Kit content
A

iCheck Fluoro Reagent Vials x 100

B

1 mL syringe x 100

C

Needle x 100

D

Vial labels/stickers

A – iCheck Fluoro Reagent Vials x 100, B – 1 mL syringe x 100, C – Needle x 100, D – Vial labels/stickers.

Technical Specifications

Sample

AnalyteAlpha tocopherol
SampleCattle serum and whole blood
Sample preparationNot necessary
Sample volume per analysis0.5 mL (500 μL)
Analytical method Autofluorescence of alpha tocopherol
Units displayed mg/L
Linear range 1.0 – 25.0 mg/L
CalibrationRequired if the value of the Standard falls outside the linear range by more than 10%
Time per analysis<10 min
Enviroment18–30°C, no direct sunlight
Method comparison High-performance liquid chromatography (HPLC)
User training 1 day training
UseLaboratory and field
Data output Sample #, Batch #, Result, Date, Time (in transferred data)
Connectivity and data Results are stored in the device and transferred to a PC via USB
Power source NiMH rechargeable batteries included; AA 1.2 or 1.5V
Warranty2 years
Device weight0.45 kg
Device dimensions11 x 4 x 20 cm (W x H x L)
Content100 reagent vials; 100 syringes – 1.0 mL; 100 needles
Chemical compositionn-Hexan and alcohols
Volume per reagent vial2.0 mL
Shelf life12 months at 20-30°C, no direct sunlight, upright
Dimension of test kit26 x 14.5 x 16.5 cm
Disposable instructionsHazardous waste
Optional equipmentManual centrifuge, 50 mL falcons, weighing dishes, reference samples

Certificates